Brief Introduction
A. Temperature displayed by LED
B. Alarm functions:
Power failure, temperature deviation, temperature sensor failure, over temperature.
C. Microprocessor based servo controlled temperature system.
D. Skin temperature sensor failure.
E. A second thermal cut-out function for more safety.
F. Adopts anti-explosion micro crystal quartz infrared radiation tube.
H. Four transparent protector board.


Technical data
Power supply
AC110/220V, 60/50Hz
Power input
1000VA
Operating condition
Ambient temperature
18°C~30°C
Ambient relative humidity
30% ~75%RH
Atmospheric pressure
700 ~1060hPa
Ambient Air velocity
<0.3m/s
Skin temperature control
32°C~38°C
Skin temperature control accuracy
≤0.5°C
Mattress temperature uniformity
<2°C
Skin temperature sensor accuracy
±0.3°C
Warm-up time
≤45 min
Mattress size
67cm (L)*54cm (W)
Distance from heater to mattress
80CM


Safety control systems
1.Adopts anti-explosion micro crystal quartz infrared radiation tube for its radiation source, the heating is quick and life-span is long.
2.Alarm functions for safety includes: Power failure, temperature deviation, temperature sensor failure, over temperature.
3.Temperature deviation can be amended in front panel directly.
4.A second thermal cut-out function for more safety.
5.Skin temperature sensor fall protect function avoid over temperature.