Поделиться:
Classifications | ||||||||
EC Directive | MDD, 93/42/EEC | |||||||
RED, 2014/53/EU | ||||||||
ROHS 2.0, 2011/65/EU | ||||||||
Degree protection against electrical shock | Type BF | |||||||
Environmental | ||||||||
Item | Operating | Storage | ||||||
Temperature | 5 to 40°C | -25 to 70°C | ||||||
Relative humidity (non-condensing) | 10% to 95% | 10% to 95% | ||||||
Barometric | 700 to 1060 hPa | 700 to 1060 hPa | ||||||
Degree of dust & water resistance | IP22 | |||||||
Drop test | 1.0 m | |||||||
Physical | ||||||||
Size (main unit) | 135mm(long)×45mm(W)×20mm(H) | |||||||
Weight (main unit) | 240 g | |||||||
Cuff size | Adult cuff: 22-42cm Small adult cuff (optional): 17-22cm Adult cuff(optional): 22-32cm | |||||||
Wireless connectivity | Built-in Bluetooth 4.0 BLE | |||||||
Power Supply | ||||||||
Charge input | Micro USB, DC5V | |||||||
Battery type | Rechargeable lithium-polymer battery | |||||||
Battery run time | 500 measurements | |||||||
Charge time | 2 hours | |||||||
Blood Pressure Measurement | ||||||||
Technology | Oscillometric Method | |||||||
Pressure measurement range | 0 – 300mmHg | |||||||
Pressure measurement accuracy | ±3mmHg | |||||||
Pulse rate range | 40 to 200 /min | |||||||
Pulse rate accuracy | ±2 /min | |||||||
Clinical accuracy | Meet IEC80601-2-30 | |||||||
ECG Recording | ||||||||
Lead type | Integrated ECG electrodes | |||||||
Lead set | Lead I, lead II, Chest Lead | |||||||
ECG length | 30s | |||||||
Heart rate range | 30 - 250/min | |||||||
Heart rate accuracy | ±2 /min or ±2%, whichever is greater | |||||||
Storage | ||||||||
Blood pressure records | 50 | |||||||
ECG records | 10 | |||||||
Bluetooth RF | ||||||||
Frequency range | 2.402 – 2.480 GHz | |||||||
Max RF power | -10 dBm |